How to Use the Touchscreen High-Current Generator Intelligent System:

Note: This intelligent control system adopts intelligent priority, with both manual and automatic control modes. When in "Automatic" mode, all operations except the emergency stop button are invalid. When in "Manual" mode, all operations except the stop button on the screen are invalid.

1. Automatic test with the touchscreen high-current generator:

(1) Connect the connecting wires according to the relevant regulations, close the power supply, and enter the main interface, as shown:


Touchscreen high-current generator interface


(2) Click on System Settings on the touchscreen high-current generator to enter the administrator password, as shown:


Touchscreen high-current generator settings

(3) Click on the dialog box and enter 8888, as shown:

Touchscreen high-current generator control system

(4) Click OK to enter the parameter setting interface, as shown:

Touchscreen high-current generator voltage and current settings

(5) Set the corresponding parameters according to the test object, then click OK:

Touchscreen high-current generator save

(6) Click Yes to return to the main interface, click Start Test to enter the test interface:

Touchscreen high-current generator test interface

(Note: The setting area is for parameters set by the user based on the test object's level, and the display area shows the data detected during the current test.)

(7) Click Start, the contactor closes, and current begins to rise, as shown:

Touchscreen high-current generator current rise section

(Note: During the voltage rise process, if you want to stop the rise, you can click the Pause button; clicking it again will continue the rise.)

(8) When the current rises to the target current, timing begins. After the specified time is reached, the system automatically returns to zero, and the test ends;

High-current generator data saving interface

(If the test object is qualified, the device will emit a beeping sound; if not, it will emit a long sound.)

(9) Click Yes on the touchscreen high-current generator to save the test data. Click Data View on the main page to enter the data saving page:

High-current generator data saving interface

2. Manual test with the touchscreen high-current generator: Note: The manual parameter settings and protection are the same as automatic. Click Start Test on the main interface, and only after entering the test interface can the manual test be performed.

(1) Close the power supply, press the Start button, the contactor closes, press the Current Rise button, and monitor the current change until it rises to the set value. At this point, the system will automatically time. After the specified time is reached, the voltage regulator automatically returns to zero, and the device emits a beeping sound, indicating that the test object is qualified.

(2) The manual timing button can be enabled at any current level.