Daily Maintenance of the Instrument


Edit


1. Within three months, if the instrument fails to meet the technical requirements due to its own reasons, it can be replaced unconditionally.


2. Free repair within one year, and lifelong maintenance and technical service.


3. If the instrument has quality problems, we will propose a maintenance plan on the same day and handle it on site promptly. We can also provide a replacement instrument to meet user needs.


4. Free on-site technical support and training on instrument usage knowledge.


5. If the instrument malfunctions, please contact us promptly so that we can provide service.


Appendix 1: Method for Replacing Paper in the Panel Printer


The printer equipped with the instrument has been installed with a paper roll at the factory. When the paper roll inside the printer is used up, a new paper roll needs to be installed. The specific installation process is as follows:


(1) Remove the front cover of the printer.


(2) Remove the entire printer from the instrument panel. Use your fingers to press the movable tabs on both sides of the printer inward, and gently take the entire printer off the instrument panel. Note: Before removing the printer, make sure the printer power is turned off.


(3) Remove the paper roll shaft from the printer.


(4) Slide the new paper roll onto the paper roll shaft, and install the shaft into the guide groove of the printer so that it does not fall out.


(5) Cut the paper head into a cone shape.


(6) Turn on the printer power, press the SEL key to turn off the SEL indicator, then press the LF key to rotate the print head. At this time, manually feed the paper head into the paper inlet under the print head. The paper will quickly enter the print head until it emerges from the front of the print head. Let it extend for a certain length, then press the SEL or LF key again to stop paper feeding. Turn off the power, close the front cover of the printer, and thread the paper head out through the slot on the front cover.


(7) Reinstall the entire printer onto the instrument panel.


Usage Method of DC Resistance Tester:


A: DC power supply test:


Close the main power switch, the corresponding indicator light will turn on. Press the "Start/Stop" key to perform the test. After the test, press the "Reset" key. The instrument will display "Discharging". Wait until the instrument returns to the startup interface, then turn off the main power switch, the corresponding indicator light will turn off. Then switch the test clips for another test.


B. AC power supply test:


Connect the AC 220V power supply, the "Charging" indicator light will turn on. Close the main power switch. The specific operation is the same as the DC power supply operation.


Special attention: After the test is completed, you must press the "Reset" key and wait until the instrument returns to the startup interface before turning off the power for the next test.