TH-2000-C Online Monitoring System for Nitrogen Oxides


Product Description:



    The TH-2000-C nitrogen oxide online monitoring system is mainly used in the gas analysis industry. Main detection applications: gas boiler exhaust monitoring, pollution source online monitoring, flue gas emission, desulfurization and denitrification, boiler exhaust, sewage pipeline gas detection and analysis, high-temperature and high-humidity gas detection and analysis, etc.



    The main principle is: the sampling unit collects the flue gas or measured gas on site, the pretreatment unit cools, dehumidifies, and filters dust from the gas, and keeps the temperature and humidity of the measured gas constant within a certain range. The gas detection unit detects and analyzes the transported gas, displays the measured gas concentration in real time on the display screen, and transmits data signals externally to terminals such as PLC or computers. It can also be transmitted via wireless GPRS (DTU) or network to local environmental protection bureaus or cloud servers, from which users can read data, enabling global networked monitoring and analysis. The entire process is automated and requires no manual intervention.



    The TH-2000-C nitrogen oxide online monitoring system is suitable for gas detection and analysis in environments with low dust, high temperature, and high humidity. If the dust content of the measured gas is high, an automatic back-flushing system or a reserved back-flushing interface for regular compressed air back-flushing is required; otherwise, regular manual maintenance and cleaning of the dust filter are needed. The TH-2000-C gas pretreatment system has a built-in two-stage electronic condensation dehumidification system and a 0.5-meter fine dust filter sampling head (flange installation), with a sampling distance of 40 meters, automatic cooling, automatic drainage, and tar filtration. For gas detection and analysis of gases that are easily soluble in water and for outdoor freezing environments in winter, to prevent freezing of the sampling pipeline, an insulation and electric heat tracing constant temperature control system is required. If high requirements are placed on the dryness and cleanliness of the measured gas, please choose the TH-3000 series, which includes an insulation and electric heat tracing constant temperature control system and a two-stage high-efficiency compressor water-cooled dehumidification system, capable of stably controlling the gas dew point at 4°C or 5°C.



    Reference regulations for the nitrogen oxide online monitoring system:



    "Emission Standard of Air Pollutants for Thermal Power Plants" GB13223-96



    "Law of the People's Republic of China on the Prevention and Control of Atmospheric Pollution" implemented from September 1, 2000



    "Technical Specification for Typical Equipment of Continuous Flue Gas Monitoring System in Thermal Power Plants" G-HB97-01



    "Technical Regulation for Continuous Emission Monitoring of Flue Gas from Thermal Power Plants" HJ/T-2000



    "Determination of Particulates and Sampling Methods of Gaseous Pollutants Emitted from Exhaust Gas of Stationary Source" GB/T16157-1996



    "Technical Conditions of Flue Gas Sampler" HJ/T47-1999



    "Technical Conditions of Flue Gas Sampler" HJ/T48-1999



    "Technical Specification for Continuous Emission Monitoring of Flue Gas from Stationary Sources" HJ/T75-2007



    "Requirements and Detection Methods for Continuous Emission Monitoring Systems of Flue Gas from Stationary Sources" HJ/T76-2007



    Product features of the nitrogen oxide online monitoring system:



    Optional explosion-proof type, waterproof design (standard is not explosion-proof, protection level IP66, outdoor type), fast, reliable, and stable detection.



    The gas analysis unit can be built into the enclosure or placed externally.



    Standard sampling distance is 30~40 meters; with optional vacuum pump, sampling distance exceeds 70 meters.



    The standard pretreatment temperature for TH-2000-C is 600°C; higher temperatures can be customized. TH3000 can pretreat flue gas at 2000°C.



    Standard: 7-inch display screen, can integrate and display various probe parameters and control other devices. Optional: no display window.



    Optional: network transmission, 4-20mA input, RS485 input, 4-20mA output, RS485 communication.



    Optional: no display window, GPRS wireless transmission, audible and visual alarm, data printing function, Linux operating system, touch screen, 16G large-capacity SD storage card, USB storage.



    Product appearance of the nitrogen oxide online monitoring system:



    Dimensions: 820mm high, 500mm wide, 350mm thick



    Principle introduction of the nitrogen oxide online monitoring system:



    The sample gas is processed according to the pressure, temperature, humidity, flow rate, (dust content), and cleanliness that the analyzer can accept, completing its processing function. It mainly completes the following tasks:



    Sample collection: Use a vacuum sampling pump or a DC brushless pump to extract the sample gas from environments such as flues.