I. Product Introduction
The Bettersize3000plus Laser Image Particle Size and Shape Analyzer is an intelligent laser particle size analyzer that uses a semiconductor-pumped 532 nm wavelength polarized laser as the light source. It adopts a single optical path system with full-angle measurement, ensuring no dead angles in scattered light detection, and offers top-level resolution, which is a patented technology of Bettersize Instruments. At the same time, it combines a dynamic particle image measurement system on the basis of laser diffraction measurement, which improves the measurement accuracy for coarse particles. Moreover, using Bettersize's proprietary technology, the laser diffraction data and image data can be fused to provide combined results, and the image method can also provide particle shape information. The combination of laser diffraction and image method is the first of its kind in China by Bettersize. Another notable feature of this instrument is its ability to measure refractive index, which is a very important parameter in laser particle size analysis. The correctness of this parameter is crucial to the accuracy of the measurement results. Therefore, Bettersize has developed an innovative refractive index measurement system based on the Bettersize3000plus instrument, combined with years of research results, ensuring the reliability and accuracy of the measurement results.
The combination of laser diffraction and image method, as well as the refractive index measurement technology, are leading-edge technologies worldwide. Therefore, the performance of the Bettersize3000plus is world-class. It also has functions such as concentration measurement, specific surface area estimation, and specific gravity estimation. Other features include automatic measurement, automatic alignment, automatic water filling, automatic drainage, automatic bubble elimination, and automatic cleaning. It uses a high-performance imported semiconductor-pumped laser with a lifespan of more than 25,000 hours; and imported high-precision lens groups for clear imaging with minimal distortion, ensuring that weak scattered light signals at various angles are not missed. The imaging system uses a German high-speed CCD and a high-quality telecentric lens, ensuring clear particle imaging. With high-speed particle recognition software, it can analyze tens of thousands to hundreds of thousands of particles per minute. This instrument has outstanding features such as accuracy, reliability, good repeatability, simple operation, and fast measurement speed. It is a new generation of high-performance particle size analyzer integrating laser technology, computer technology, digital image technology, and optoelectronic technology. It is at the world's leading level and can compete with foreign particle size analyzers. It is especially suitable for laboratories in enterprises, universities, and research institutes.
II. Main Technical Indicators and Performance
-
Laser Diffraction System
| Measurement Range | 0.01-3500μm | Sample Feeding Method | Automatic Circulation Dispersion System |
| Repeatability Error | ≤0.5% (D50 deviation of national standard sample) | Accuracy Error | ≤0.5% (D50 deviation of national standard sample) |
| Bimodal Resolution | ≥1.65 (ratio of nominal values of two standard samples) | Measurement Principle | Mie Scattering Theory |
| Measurement Mode | Fully Automatic Measurement (SOP) | Fastest Measurement Speed | ≤10 seconds |
| Laser Light Source | Polarized Pumped Solid-State Laser (5mW/532nm) | Optical Path System | Oblique Incidence Dual Lens, Automatic Alignment |
| Dew Point Temperature Measurement | Yes | Detection Angle | 0.02-165o |
| Refractive Index Measurement Range | 1.4-3.6 | Operating System | WinXP/Win7/Win8/Win10 |
| Interface | USB2.0 or 3.0 | Photodetector | 96 elements |
| Ultrasonic Power | 50W | Circulation Cell Volume | 600ml |
| Circulation Flow Rate | 3000-8000ml/min | Product Compounding | X product + Y product = Z product |
| Voltage | AC220V, 50/60Hz | Dimensions, Weight | 820×610×290mm, 48kg |
-
Microscopic Image System
| Image Analysis Range | 2-3500μm | Imported High-Speed CCD | 120 frames/second |
| Telecentric Imaging System | Magnification 20-100 times | Image Recognition Speed | 10,000 particles/minute |
| Analysis Items | Particle size distribution, maximum particle size, aspect ratio, circularity, etc. | ||
III. Main Application Fields
-
Various non-metallic powders: such as ground calcium carbonate, precipitated calcium carbonate, talc, kaolin, graphite, wollastonite, brucite, barite, mica powder, bentonite, diatomite, clay, silica, garnet, zirconium silicate, zirconia, magnesium oxide, zinc oxide, etc. -
Various metal powders: such as aluminum powder, zinc powder, molybdenum powder, tungsten powder, magnesium powder, copper powder, as well as rare earth metal powders, alloy powders, etc. -
Other powders: river sediment, lithium battery materials, catalysts, phosphors, cement, abrasives, pharmaceuticals, pesticides, food, coatings, dyes, ceramic raw materials, chemical materials, nanomaterials, papermaking fillers and coatings, various emulsions, etc.
IV. Outstanding Features
-
Oblique Incidence Dual Lens Technology: Bettersize patented technology, using oblique laser incidence combined with forward, side, and backscattered light detection to achieve full-angle measurement, expanding the measurement range, improving the measurement accuracy of fine particles, and enhancing resolution. -
Microscopic Image and Laser Diffraction Two-in-One Technology: The laser diffraction method has advantages in measuring fine particles, while the image method has high accuracy for coarse particles. The combined laser and image testing is a strong alliance, resulting in higher accuracy of test results, and also enables particle shape analysis. -
Sample Refractive Index Measurement Technology: For samples with unknown refractive index, the refractive index can be measured first, including the real and imaginary parts, ensuring the accuracy of particle size measurement. -
Automatic Circulation Dispersion and Automatic Measurement Technology: Anti-dry-burning ultrasonic disperser, centrifugal circulation pump, automatic water filling system, automatic drainage and overflow system, suitable for all samples, ensuring full dispersion of samples, and ensuring accuracy and repeatability of tests.
V. Good Repeatability
-
Stable and effective dispersion system. -
The laser, detector, and signal transmission system are stable and reliable. -
The automatic alignment system keeps the instrument in an ideal state at all times. -
The sampling speed reaches 3500 times/second, and a large amount of data effectively reduces the impact of a few abnormal data on repeatability.
(Repeatability of continuous testing of the same sample)