Transistor Characteristic Curve Tracer
System Block Diagram
The computer is installed with a PCI interface board and BJ4822 operating software; the computer controls the BJ4822 test host through the PCI bus; simultaneously reads the data in the acquisition memory of the BJ4822 test host; the characteristic curves of semiconductor devices are displayed through the BJ4822 operating software. Point marking test is provided. The device under test can be connected to the test box via a test adapter, or directly connected to the test box via test leads; each source is connected to the panel of the test box; the user selects different sources to inject into the various pins of the device under test as needed.
Main Technical Specifications
(1) Collector Pulse Source (High Current Source):
Short-circuit current: 3kW—maximum short-circuit current 400A ;
300W—maximum short-circuit current 40A ;
Open-circuit voltage: 30V (+10%, -5%)
Adjustable range: continuously adjustable from 0% to 100%, resolution 0.5%
Pulse width: 300μs±10%
Repetition frequency: 3kW—1/4 power frequency; 300W—1/2 power frequency
(2) Collector Sweep Source (High Voltage Source):
Short-circuit current: 30W—40mA; 3W—4mA;
Open-circuit voltage: 3kV (+10%, -0%)
Adjustable range: continuously adjustable from 0% to 100%, resolution 0.1%
(3) Step Voltage Source:
Step amplitude range: 200mV/step—5V/step (in 1-2-5 sequence); 5 ranges in total; error of each range ≤±3%
Number of steps: 0—5 steps continuously variable
(4) Step Current Pulse Source:
Step range: 1mA/step— 2A /step (in 1-2-5 sequence); 11 ranges in total; error of each range ≤±3%
Pulse width: 550μs±10%
Number of steps: 0—5 steps continuously variable
(5) Step Current Source:
Step range: 1μA/step—2mA/step (in 1-2-5 sequence); 11 ranges in total; error of each range ≤±3%
Number of steps: 0—5 steps continuously variable
(6) Horizontal Voltage Measurement:
(a) Vce voltage source setting (in 1-2-5 sequence):
3kW, 300W: 100mV/div—5V/div (in 1-2-5 sequence); 5 ranges in total; error of each range ≤±3%
30W, 3W: 50V/div—500V/div (in 1-2-5 sequence); 4 ranges in total; error of each range ≤±5%
(b) Vbe voltage source
100mV/div—5V/div (in 1-2-5 sequence); 5 ranges in total; error of each range ≤±3%
(7) Collector Current Measurement:
Vertical range (in 1-2-5 sequence):